MODEL 8559A

PERFORMANCE TESTS

PERFORMANCE TESTS

4-25. CALIBRATOR OUTPUT ACCURACY

SPECIFICATION:

Amplitude: - 10 dBm +0.3 dB

Frequency: 35 MHz +400 kHz

DESCRIPTION:

The frequency of the calibrator output signal is measured with a microwave counter. The calibrator output level is measured using a power meter.

SPECTRUM

ANALYZER

FREQUENCY

POWER

COUNTER

METER

 

 

 

 

0 y:;;

 

 

 

 

@.a@@

 

1 - I

0..

0...

0

-

 

 

I INPUT

 

I OUTPUT

 

POWER

 

 

 

 

 

SENSOR

FIGURE 4-21,CALIBRATOR ACCURACY TEST SETUP

EQUIPMENT:

 

 

 

 

 

Frequency Counter ..............................................

 

 

 

H P 5342A, Opt. 005

Power Meter ...........................................................

 

 

 

H P 435A/B

Power Sensor ...........................................................

 

 

 

H P 8481A

PROCEDURE:

1.Connect spectrum analyzer CAL OUTPUT to frequency counter's 10 Hz - 500 MHz (5m) input as shown in Figure 4-20. Measured output frequency should be 35 MHz k400 kHz.

2.Zero and calibrate power meter. Connect power sensor, through adapter, to spectrum analyzer's CAL OUTPUT and measure power level. Calibrator output level should be - 10 dBm +0.3 dB.