MODEL 8559APERFORMANCETESTS

PERFORMANCE TESTS

 

4-22. INPUT ATTENUATOR ACCURACY(Cont'd)

 

PROCEDURE:

 

Connect equipment as shown in Figure 4-17and set controls as follows:

 

Spectrum Analyzer:

 

FREQUENCY BAND GHz ..................................................

.01- 3

TUNING .................................................................

30MHz

FREQSPAN/DIV ..........................................................

2MHz

RESOLUTIONBW .........................................................

lMHz

INPUTATTEN .............................................................

60dB

REFERENCELEVEL .......................................................

OdBm

REFLEVELFINE ..............................................................

0

Amplitudescale ........................................................

10dB/DIV

SWEEPTIME/DIV ........................................................

AUTO

SWEEPTRIGGER .....................................................

FREERUN

ALTIF .....................................................................

OFF

SIGIDENT .................................................................

OFF

BLCLIP ....................................................................

OFF

VIDEOFILTER ..........................................................

2o'clock

Signal Generator:

 

COUNTER MODE ..............................................

INT EXPAND XlO

AM .......................................................................

OFF

FM ........................................................................

OFF

FREQUENCYTUNE ....................................................

30.0MHz

RF ..........................................................................

ON

OUTPUTLEVEL ..........................................................

OdBm

Set step attenuator to 0 dB and use spectrum analyzer TUNING control to center 30 MHz signal from signal generator on CRT display. Set FREQ SPAN/DIV to 20 kHz, RESOLUTION BW to 10 kHz, and Amplitude Scale to 1 dB/DIV.

Adjust signal generator OUTPUT LEVEL control to position peak of signal seven divisions above graticule baseline.

Set step attenuator and INPUT A m E N control to settings indicated in Table 4-9.For each setting, record deviation of signal peak from referencegraticule line set in step 3.

NOTE

The REFERENCE LEVEL control setting changes by 10 dB for every 10-dB change in INPUT ATTEN. Do not change the REFERENCE LEVEL setting after changing the INPUT ATTEN setting.