MODEL 8559A

PERFORMANCETESTS

PERFORMANCE TESTS

4-23. REFERENCE LEVEL ACCURACY

SPECIFICATION:

Step Accuracy (steps referenced with 0 dB input attenuation):

-10 dBm to - 80 dBm: +0.5 dB

-10dBmto-100dBm: +1.0dB

Vernier Accuracy: f0.5 dB

DESCRIPTION:

The reference level accuracy is tested over the range of - 10 dBm to - 100 dBm by checking the IF gain steps in 1 dB/DIV (Log) and in LIN. The resulting maximum deviation in each case must be less than k0.5 dB from

-10 dBm to - 80 dBm and less than k 1.0 dB from -10 dBm to - 100 dBm.

SPECTRUM

ANALYZER

SIQNAL QENERATOR

STEP

ATTENUATOR

3 5 5 0

 

OPT H82

 

ADAPTER t

A!!

 

 

STEP

 

 

ATTENUATOR

 

 

S55C

 

 

OPT H8O

 

FIGURE4-18.REFERENCE LEVELACCURACYTEST SETUP

 

EQUIPMENT:

 

 

SignalGenerator .........................................................

HP8640B

10-dB Attenuator ................................................

H P 8491B, Opt. 010

Step Attenuator (1 dB/step) ........................................

H P 355C, Opt. H80

Step Attenuator (10 dB/step) ......................................

H P 355D, Opt. H82

Adapter, Type N (m) to BNC (f) (2 required) ...............................

H P 1250-0780