PERFORMANCETESTS

MODEL8559A

PERFORMANCE TESTS

4.22.INPUT ATTENUATOR ACCURACY (Cont'd)

TABLE 4-9.INPUT AllENUATOR ACCURACY

INPUT ATTEN

Step

 

 

Amplitude

Step Attenuator

 

Corrected

 

Setting

Attenuator

 

 

Deviation

 

Error

 

Deviation

 

(dB)

Setting (dB)

 

 

(dB)

(Calibration)"

 

(dB)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

I

*Attenuations > dial settings are positive (+). Attenuations <dial settings are negative (-).For example, 9.99 dB

I

calibration for a 10 dB attenuator setting represents an error of -0.01 dB.

To compute corrected deviation for each setting, add step attenuator error to amplitude deviation. Corrected deviation should not exceed 5 1.0 dB between any two adjacent INPUT ATTEN setting.

dB Maximum Error per 10-dB Step

Record maximum positive and maximum negative corrected deviation values. Difference between these two values (maximum cumulative step error) should not exceed 2.4 dB.