ADJUSTMENTS

MODEL 8559A

ADJUSTMENTS

5-30. FREQUENCY DISPLAY ADJUSTMENTS(Cont'd)

PROCEDURE:

Allow one-half hour warmup time of equipment with analyzer connected to mainframe with extender cable.

Jumper A8TP5 DPM to ground.

Set front-panel FREQUENCY BAND GHz to Band 1 (.01- 3).

Connect DVM to AlA2TPl DPMA. AlA2TPl is located below the board and is accessible through cutout in left side gusset.

Adjust A8R61 DPM ZERO for DVM indicationof 0.000 Vdc.

Adjust AlA2R29 OFFSET for front-panel FREQUENCY GHz indicationof 0.000.

Remove jumper from A8TP5 to ground.

Select Band 6 (12.1 - 21) on analyzer.

Adjust front-panel TUNING control for DVM indicationof - 4.000 Vdc.

Adjust AlA2R28 GAIN for front-panel FREQUENCYindication of 20.000.