ADJUSTMENTS

MODEL 8559A

ADJUSTMENTS

 

5-19. LOG AMPLIFIER LOG AND LINEAR ADJUSTMENTS(Cont'd)

 

PROCEDURE:

 

1. Set equipment controls as follows:

 

Spectrum Analyzer:

 

FREQUENCY BAND GHz ..................................................

.01 - 3

TUNING .............................................................

>O.OlOGHz

FREQSPAN/DIV ...............................................................

0

RESOLUTION BW ..............................................

300 kHz, uncoupled

INPUTATTEN .............................................................

10dB

REFERENCELEVEL ....................................................

-50dBm

REFLEVELFINE .............................................................

0

Amplitude Scale ..............................................................

LIN

SWEEP TIME/DIV ........................................................

AUTO

SWEEPTRIGGER .....................................................

FREERUN

ALTIF .....................................................................

OFF

SIG IDENT .................................................................

OFF

BLCLIP ....................................................................

OFF

VIDEOFILTER ............................................................

OFF

Signal Generator:

 

COUNTERMODE ...........................................................

INT

AM ........................................................................

OFF

FM ........................................................................

OFF

FREQUENCY TUNE ...................................................

321.4 MHz

RF ..........................................................................

ON

OUTPUTLEVEL ................................................

approx. -28dBm

2.Set 1-dB step attenuator to 10 dB and 10-dB step attenuator to 0 dB. Remove AlOWl (blue cable) from A5J2 and connect equipment as shown in Figure 5-3, using adapter to connect step attenuator to AlOWl.

NOTE

The HP 355C 10 dB attenuation is included to compensate for 10 dB of gain on Step Gain Assembly A12 with the TEST-NORM switch in TEST.

3.Set TEST-NORM switch on Step Gain Assembly A12 to TEST position. Adjust signal generator FRE- QUENCY TUNE control for maximum signal amplitude on display with 10-dBstep attenuator set to 0dB (reduce signal generator OUTPUT LEVEL control setting as necessary to bring signal on-screen).