ADJUSTMENTS

MODEL 8559A

ADJUSTMENTS

5-29. CAL OUTPUT AND REF LEVEL CAL ADJUSTMENTS(Cont'd)

15.Signal peak should now be approximately at top graticule line (Reference Level) on CRT.

16.Switch between 10 dB/DIV and LIN while adjusting A12R57 1A (offset) to place signal peak at same level in both 10 dB/DIV and LIN.

17.Level at which signal peaks are coincident should be at top graticule line (Reference Level). If not, adjust front-panel VERTICAL GAIN to place signal peak at Reference Level line. Be sure VERTICAL POSN is properly adjusted for baseline on bottom graticule line.

18.Replace CAUTION label (PC Board) on A12 Step Gain.