PERFORMANCETESTS

MODEL 8559A

 

PERFORMANCE TESTS

--

4-19.FREQUENCY RESPONSE (Cont'd)

21.Select 9 - 15 GHz frequency band on spectrum analyzer. Set sweep oscillator to CW and frequency to 12.0 GHz. Use CW control to center signal on spectrum analyzer display. Trigger a 6 GHz sweep. Adjust spectrum analyzer TUNING control to reposition tuning marker. Trigger another sweep. Record greatest positive and greatest negative deviation of signal peaks from center horizontal graticule line (neglect devia- tions caused by tuning marker).

Maximumdivisions

Minimumdivisions

22.Subtract mean deviation of step 16 from maximum and minimum values recorded in step 21. Frequency response for 9 - 15 GHz frequency band should not exceed +1.8 dB (+1.8 divisions).

Maximum

 

 

divisions (9- 15 GHz)

Minimum

 

 

divisions (9- 15 GHz)

23.Select 6 - 15 GHz frequency band on spectrum analyzer. Set sweep oscillator to CW and frequency to 10.5 GHz. Use CW control to center signal on spectrum analyzer display. Set AF control for 9 GHz and trigger a sweep. Adjust spectrum analyzer TUNING control several turns to reposition tuning marker. Trigger another sweep. Record greatest positive and greatest negative deviation of signal peaks from center hori- zontal graticule line (neglect deviations caused by tuning marker).

Maximumdivisions

Minimumdivisions

23.Subtract mean deviation of step 16 from maximum and minimum values recorded in step 23. Frequency response for 6 -15GHz frequency band should not exceed k2.1 dB (+2.1 divisions).

Maximum

 

divisions (6 - 15 GHz)

Minimum

 

divisions (6 -15GHz)

25.Select 12.1 -21GHz frequency band on spectrum analyzer and adjust TUNING control fully clockwise. Set sweep oscillator to CW and frequency to 15 GHz. Set AF control for 6 GHz and trigger a sweep. Record greatest positive and greatest negative deviation of signal peaks from center horizontal graticule line (12.1 GHz to 18 GHz).

Maximumdivisions

Minimumdivisions