ADJUSTMENTS

MODEL 8559A

 

ADJUSTMENTS

 

5-28. FREQUENCY RESPONSE ADJUSTMENTS(Cont'd)

 

5 .

Set spectrum analyzer controls as follows:

 

 

REF LEVEL FINE ..........................................................

0 dBm

 

REFLEVELdBm ............................................................

- 20

 

INPUTATTEN .............................................................

10dB

 

Amplitude Scale ........................................................

10 dB/DIV

 

FREQ SPAN/DIV ..........................................................

F (full)

 

RESOLUTIONBW .........................................................

3MH z

 

FREQUENCYBANDGHz ..................................................

.01-3

 

TIME/DIV ...............................................................

20msec

 

TRIGGER ............................................................

FREERUN

 

BLCLIP ....................................................................

OFF

 

VIDEOFILTER .............................................................

MIN

 

ALTIF .....................................................................

OFF

 

SIGIDENT .................................................................

OFF

 

TUNING ................................................................

2.5GHz

.01- 3 GHz Adjustment

 

6 . Place sweep oscillator plug-in RF OFF/ON switch to ON.

 

7 .

Adjust sweep oscillator controls for a CW output of 2 GHz at - 7 dBm.

 

8.Disconnect power meter/power sensor and connect 20 dB attenuator directly t o analyzer RF INPUT as shown in Figure 5- 16.

9.Adjust sweep oscillator controls for full sweep.

10.Adjust Tuning Voltage Circuit GAIN control fully clockwise then adjust OFFSET control to center phase- locked signal on CRT. Refer to Figure 5-18a.

FIGURE 5-18. ILLUSTRATION OF PHASE-LOCKING PROCEDURE