ADJUSTMENTS

MODEL 8559A

ADJUSTMENTS

5-26. SECOND CONVERTER ADJUSTMENTS(Cont'd)

PROCEDURE:

Second LO PreliminaryAdjustment

Allow one-half hour warm-up time of equipment with analyzer connected to mainframe with extender cable.

Connect frequency counter input to A5J3 2nd LO output using the test cable and two SMB adapters.

Select Band 1 (.01-3) and Alternate IF on spectrum analyzer front-panel by depressing these pushbut- tons.

Using the special Allen driverhut driver tuning tool, adjust A5Z4 2nd LO FREQUENCY for a frequency counter indication of 2671.1 MHz k0.5 MHz.

Connect spectrum analyzer CAL OUTPUT to RF INPUT and adjust front-panel TUNING controls to center the calibrator signal on the CRT display.

Set spectrum analyzer controls as follows:

 

FREQ SPANiDIV ..........................................................

1MHz

RES BW .................................................................

300 kHz

TIME/DIV ...............................................................

2mSEC

FREQUENCYBANDGHz .........................................

BANDl(.Ol-3)

SIG IDENT .................................................................

OFF

ALTIF .....................................................................

OFF

853A ............................

TRACE A & B STORE BLANK (ANALOG DISPLAY)

Depress front-panel SIG IDENT and ALT IF pushbuttons.

Turn SIG IDENT off and on while monitoring the display. The signal traces which appear when SIG IDENT is switched on are the signal identifier signals. The others are the alternate IF signals.

Adjust TUNING to place one of the signal identifier signals on a graticule line. This will be the reference graticule line.

Turn ALT IF off. Adjust A8R34 REG to center the signal identifier signal on the reference graticule line.

Turn ALT IF on. Verify that the signal identifier signal appears on reference graticule line. If not, repeat step 10.

Adjust A8R40 SIG ID to place the alternate IF signal 1 MHz (1 division) higher than the signal identifier signal.

Turn ALT IF off. Adjust A8R39 OFF to center the signal on the same graticule line as the alternate IF signal (1 MHz higher than reference graticule line).