GENERAL INFORMATION

 

 

 

MODEL 8559A

 

 

TABLE 14. RECOMMENDED TEST EQUIPMENTQOF 4)

 

 

 

 

 

 

 

 

 

 

 

Instrument

Critical Specifications

 

Recommended

Use*

 

 

Model

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Fixed Attenuator

Frequency Range: 10 MHz t o 18 GHz

 

HP 8491B

 

 

 

(2 required)

Attenuation: 20 dB + I .O dB

 

Option 0206

 

 

 

 

Connectors: Type N (m) ( f )

 

 

 

 

 

Fixed Attenuator

Frequency Range: 10 MHz t o 18 GHz

 

HP 8491B

 

 

 

 

Attenuation: 10 dB +0.5 dB

 

Option 0106

 

 

 

 

Connectors: Type N (m) ( f )

 

 

 

 

 

Crystal Detector

Frequency Range: 10 MHz to 21 GHz

 

HP 33330C

 

 

 

 

Frequency Response: t 0 . 6 dB, .O1 to 18 GHz

 

 

 

 

 

 

Maximum SWR: G1.5, .O1 to 18 GHz

 

 

 

 

 

 

Output Polarity: Negative

 

 

 

 

 

 

Connectors:

 

 

 

 

 

 

APC -3.5 (SMR) (m) input, SMC (m) output

 

 

 

 

 

Termination

Frequency Range: 1 0 MHz to 18 GHz

 

HP 909A

 

 

 

 

Impedance: 5052

 

Option 0 1 2 ~

 

 

 

 

Connector: Type N (m)

 

 

 

 

 

Tuning Voltage Circuit

Refer to Figure 5-17

 

None

 

 

 

Crystal Bypass Networks

Refer to Figure 5-6

 

None

 

 

 

Special Extender Board

Refer to Figure 5-9

 

None

 

 

 

Extender Cable

Extends Spectrum Analyzer Plug-In for Servicing

 

HP 5060-0303

 

 

 

 

Refer to Figure 1-3.

 

 

 

 

 

Cable

Frequency Range: 10 MHz to 21 GHz

 

 

 

 

 

 

Maximum SWR: G1.4 at 21 GHz

 

 

 

 

 

 

Length: 6 1 cm (24 inches)

 

 

 

 

 

 

Connectors: SMA (m) both ends

 

 

 

 

 

Cable

BNC (m) to SMC (f), 3 6 inches long

 

 

 

 

 

Cable

48 inch, 50!2 coaxial cable with BNC (m)

 

 

 

 

 

 

connectors on both ends (3 required)

 

 

 

 

 

Cable

RG-214/U with Type N connectors (2 required)

 

 

 

 

 

Cable

BNC (m) to Banana Plug

 

 

 

 

 

Test Cable

Connectors: BNC (m) t o SMB (f)

 

 

 

 

 

 

Length: >61 cm (24 inches)

 

 

 

 

 

 

 

 

 

 

 

*P = Performance Test; A = Adjustments; T = Troubleshooting Option number specifies attenuation value.

Option 012 provides type N male connector.