PERFORMANCE TESTS

MODEL8559A

PERFORMANCE TESTS

4-11.FREQUENCY SPAN ACCURACY (Cont'd)

13.Set function generator frequency to 200 kHz (k0.5%) using frequency counter. Connect function genera- tor output to comb generator MODULATION input. Set function generator OUTPUT LEVEL for a clean 200-kHz comb on the spectrum analyzer display.

NOTE

To obtain a clean comb on the spectrum analyzer display, use either the LOW or HIGH output of the function generator and readjust the OUTPUT LEVEL control as necessary.

14.Set spectrum analyzer FREQ SPAN/DIV to 200 kHz. Adjust TUNING control to position one spectral line on the first graticule line. Measure the error between ninth spectral line and ninth graticule line. Error should be no greater than k0.4division.

div

15.Using the procedure of NOTE in step 13,vary spectrum analyzer FREQ SPAWDIV and function genera- tor output frequency in accordance with Table 4-2. Adjust spectrum analyzer TUNING control to position one spectral line on the first graticule line. Measure the span error between ninth spectral line and ninth graticule line.

NOTE

Disconnect function generator from comb generator when setting frequency with frequency counter. Increase spectrum analyzer REFERENCE LEVEL control setting as necessary for the lowest frequencies.

TABLE 4-2.NARROW SPAN WIDTH ERROR MEASUREMENT

I

 

 

 

 

 

I

Spectrum Analyzer

 

Span Width Error

 

Function Generator

 

 

 

 

 

 

FREQ SPANIDIV

RESOLUTION BW

Output Frequency*

 

 

 

 

 

 

 

 

 

 

 

 

 

100 kHz

OPTIMUM

100 kHz

k0.4 div.

-div.

 

50 kHz

OPTIMUM

50 kHz

+0.4 div.

-div.

 

20 kHz

OPTIMUM

20 kHz

k0.4 div.

-div.

 

10kHz

OPTIMUM

10kHz

k0.4 div.

-div.

 

 

 

 

 

 

*Check function generator output frequency using a frequency counter. Frequency readout should be within f0.5% of desired audio frequency.