MODEL 8559A

ADJUSTMENTS

ADJUSTMENTS

5-17. POWER SUPPLY CHECKS AND ADJUSTMENTS(Cont'd)

NOTE

The two following voltage adjustments, + 10V and VO (Varactor Offset), must be performed while the spectrum analyzer is still cold (during first five minutes after turn-on). If the instrument has been operating longer than five minutes, turn off the display mainframe, remove A8 and A9 assemblies, and let them cool on bench for 15 minutes. Replace the two assemblies and proceed with adjustment of A9R2 and A8R62 during the first five minutes after turn-on.

7. Connect digital voltmeter to + 10V test point A9TP7 and adjust + 10V potentiometer A9R2 for a voltmeter indication of + 10.000 +0.100 Vdc.

8.Connect digital voltmeter to VO test point A8TP2. Set spectrum analyzer SWEEP TIME/DIV control to 10 ms and SWEEP TRIGGER control to SINGLE. Turn ALT IF and SIG IDENT on (pushbuttons depressed).

9.Voltage at A8TP2 will alternate between two values each time a sweep is triggered. Trigger sweep a few times until voltmeter indicates least negative VO voltage. Adjust VO potentiometer A8R62 for a voltmeter indication of - 2.00 & 0.10 Vdc.