MODEL 8559A

ADJUSTMENTS

ADJUSTMENTS

-

5-17. POWER SUPPLY CHECKS AND ADJUSTMENTS

REFERENCE:

A7, A8, A9 Schematics

DESCRIPTION:

The + 14.5V and - 10V regulated power supplies on Frequency Control Assembly A7 are adjusted. The (dependent) - 12V power supply is then checked for proper dc output (with less than +50 mV variation) while the spectrum analyzer is tuned from 10 MHz to 3 GHz. The + 10V power supply on Sweep Generator/Band- width Control Assembly A9 and the VO (Varactor Offset) voltage on Marker Assembly A8 are then adjusted. Both the + 10V power supply voltage and the VO voltage are temperature-dependent and must be adjusted during the first five minutes after the spectrum analyzer is turned on (cold instrument).

D I G I T A L VOLTMETER

D l SPLAY

CABLE ASSEMBLY

FIGURE 5-1. POWER SUPPLYCHECKS AND ADJUSTMENTSTESTSETUP

EQUIPMENT:

 

Digital Voltmeter ........................................................

HP 3456A

Extender Cable .......................................................

HP 5060-0303

Cable Assembly, BNC (m) to Banana Plug ...................................

HP 11001A

Adapter, BNC (f) to Alligator Clips ......................................

HP 8120-1292