SERVICE

MODEL8559A

Resistance R, is at a minimum (approximately 150 ohms) for small signals. The small signal gain of the stage (about 10 dB) is established by the dc bias through the log diodes. As the signal level at the emitter of 413 increases, signal current cancels bias current in the log diodes, increasing R,. The gain of the stage for large signals is reduced to unity (0 dB) as R, becomes very large.

Linear Mode of Operation. Two simplified schematics illustrating unity and 10 dB gain of a typical linear stage are shown in Figures 8-65 and 8-66. In linear mode, the signal level dependent components are removed from the signal path and a linear display is provided. The - 8 VT is applied to the base of 424, turning it off. This removes dc bias from CRlO and CRll . Total resistance R, (primarily the resistance of R56 and CR12) is high, since CR12 is reverse-biased. Control line IFG6 is high and the stage gain is near unity. The signal flow is through emitter follower 413 and R52, to 420. In stages six and seven, an alternate signal path is used to fix the gain at about 5 dB per stage, allowing for scale differences between Log and Lin modes. Both stages are activated by the - 8 VT from the Amplitude Scale switch through R34, R93, R101, CR25, and CR28. The combined stage gain is adjusted by R34 (LIN), which controls the dc PIN diode bias.

/

 

SECOND STAGE

 

\

OUT

LOGILI N LIN E - 8 V T @-$

w4

FIGURE 8-65. UNITYGAIN OPERATION IN LINEAR MODE, SlMPLlFlEDSCHEMATlC

Stage 2, 3,4, and 5 each have an alternate signal path that switches in 10 dB of step gain for a total of 40 dB. The alternate path is selected by the REFERENCE LEVEL control. With the INPUT ATTEN at 0 dB and the REFERENCE LEVEL control at -68dBm, the - 8 VT is routed, via the IF gain control line (IFG4), to forward bias CR22 in stage 5. For each stepped increase in the REFERENCE LEVEL control, the - 8 VT activates the IFG lines associated with the stages of gain required, forward biasing the diodes in the signal path. Each IFG line has a potentiometer (block B) that controls the line's bias current and the stage gain. Note that IFG6 controls two stages (stages 2 and 3) that, when switched in, provide 20 dB of gain.

Gain Control Lines (B)

The + 15V (in Log mode) or the - 8 VT (in Lin mode) is routed through the REFERENCE LEVEL switch to the combination of IFG4, IFG5, and IFG6 corresponding to the reference level selected. In Log mode, the Log Offset circuit is activated through R24,R25, and R26. The LOG/LIN line is at + 15V,424 is saturated, and the

0-168