ADJUSTMENTS

5-21.BANDWIDTH FILTER ADJUSTMENTS(SERIAL PREFIX 1909A)(Cont'd)

16.Adjust A13C15 SYM and A13C25 CTR for a centered and symmetrical bandpass. Adjust A13C25 CTR for minimum signal amplitude,

17.Remove crystal short from A1lTP4/TPS..

18.Adjust A11C38 SYM and AllC54 CTR for a centered and symmetrical bandpass. Adjust AllC54 for minimum signal amplitude.

19.Remove crystal short from A11TP1/TP2.

20.Adjust AllC15 SYM and AllC25 CTR for a centered and symmetrical bandpass. Adjust AllC25 for minimum signal amplitude.

21.Remove the crystal shorts.

LC Alignment

22.Perform preliminary LC filter adjustmentsas follows:

NOTE

When Bandwidth Filter Assemblies A11 and A13 are installed with coversin place, midget copper alligator clips (HP Part Number 1400-0483)can be used to short test points to the cover.

a.Install A13 on extender board.

b.Short to ground the following test points: A13TP6, AllTP3, and AllTP6. Jumper A9TPl to A9TP2.

c.Adjust A13C73 lor minimum signal amplitude.

d.Disconnect short from A13TP6and short to ground A13TP3.

e.Adjust A13C74 for minimum signal amplitude.

f.Reinstall A13 and install A11on extender board.

g.Disconnect short from A13TP3and short to ground A11TP6.

h.Adjust A11C73for minimum signal amplitude.

i.Disconnect short from A11TP6and short to ground A1 1TP3.

j.Adjust A11C74for minimum signal amplitude.

k.Disconnect shorts from test points and reinstall A11. Replace covers on A11 and A13 assemblies. Remove jumper from A9TP1/A9TP2.