ADJUSTMENTS

5-21. BANDWIDTHFILTER ADJUSTMENTS(SERIAL PREFIX 1909A)(Cont'd)

23.Carefully center signal on CRT in 30 kHz RESOLUTION BW; then switch RESOLUTION BW to 100 kHz. Note wheresignal intersects the center vertical graticule line.

24.Adjust A13C45 LC CTR for maximum signal amplitude where the signal intersects the center vertical graticule line.

25.Switch RESOLUTION BW to 30 kHz and center signal; then switch to 100 kHz. Note where signal intersects the center vertical graticule line.

26.Adjust A13C23 LC CTR for maximum signal amplitude where the signal intersects the center vertical graticule line.

27.Switch RESOLUTION BW to 30 kHz and center signal; then switch to 100 kHz. Note where signal intersects the center vertical graticule line.

28.Adjust A1lC45 LC CTR for maximum signal where the signal intersects the center vertical graticule line.

29.Switch RESOLUTION BW to 30 kHz and center signal; then switch to 100 kHz. Note where signal intersects the center vertical graticule line.

30.Adjust AllC23 LC CTR for maximum signal amplitude where the signal intersects the center vertical graticule line.

3l. Switch RESOLUTION BW between 100 kHz and 30 kHz to be sure the signal is centered at both bandwidth settings.

Bandwidth Amplitude

32.Set Amplitude Scale to 1 dB/DIV and SWEEP TIMWDIV to AUTO.

33.Set RESOLUTIONBW to 3 MHz and FREQ SPAN/DIV to 50 kHz.

34.Adjust fine TUNING and REF LEVEL FINE for a centered signal at 7 divisions.

35.Set RESOLUTION BW to 100 kHz and center signal with fine TUNING control. Adjust A13R26 LC and A11R26LC equally to obtain a signal amplitudeof 7 divisions.

36.Set RESOLUTION BW to 1 kHz and FREQ SPAN/DIV to 10 kHz. Center signal with fine TUNING control. Adjust A11R31XTL and A13R31 XTL equally for a signal amplitudeof 7 divisions.

NOTE

Each potentiometer should be adjusted to accomplish half the necessary increase in signalamplitude.