MODEL 8559A

ADJUSTMENTS

ADJUSTMENTS

5-21. BANDWIDTH FILTER ADJUSTMENTS(Cont'd)

TERMINAL CONNECTORS

FIGURE 5-6.CRYSTALSHORTCONFIGURATION

 

PROCEDURE:

 

NOTE

 

Allow 30 minutes warmup time before performingadjustments.

 

1 . Set spectrum analyzer controls as follows:

 

FREQUENCYBAND ..................................................

.Ol-3GHz

TUNING .................................................................

35MHz

FREQSPAN/DIV .........................................................

200kHz

RESOLUTIONBW .........................................................

lMHz

INPUTATTEN .............................................................

10dB

REFERENCELEVEL .......................................................

OdBm

Amplitude Scale ..............................................................

LIN

SWEEPTIME/DIV ........................................................

10msec

SWEEPTRIGGER .....................................................

FREERUN

Crystal Alignment

2.Connect equipment as shown in Figure 5-5.

NOTE

If Sweep Generator A9 has been replaced or adjusted, perform steps 3 through 8. If not, proceed to step 9.