MODEL 8559A

ADJUSTMENTS

ADJUSTMENTS

5-22. 3.dB BANDWIDTH ADJUSTMENTS

REFERENCE:

A9 Schematic

DESCRIPTION:

The 3-dB bandwidths for the 3 MHz through the 30 kHz RESOLUTION BW settings are adjusted using the CAL OUTPUT as the signal source. The 3-dB bandwidths for the 10 kHz, 3 kHz, and 1 kHz RESOLUTION BW settings are adjusted by injecting a stable 321.4 MHz signal into the Third Converter (A10) of the spectrum analyzer.

SIDNAL DENERATOR

MICROWAVE

COUNTER

D l SPLAY

 

 

 

l NPUT

SPECTRUM

 

 

 

P

ANALYZER

 

 

 

m 0

 

ADAPTER

ADAPTER

 

 

 

OUTPUT

ADAPTER !EE

ATTENUATORSTEP

 

FIGURE5-8. 3-dBBANDWIDTH ADJUSTMENTTESTSETUP

 

 

EQUIPMENT:

 

 

 

 

Signal Generator .........................................................

 

 

HP 8640B

Frequency Counter ..........................................................

 

 

5342A

StepAttenuator ..........................................................

 

 

HP355D

Adapter, Type N (m) to BNC (f) (2 required) ...............................

 

HP 1250-0780

Extender Cable Assembly ..............................................

 

 

HP 5060-0303

Test Cable, BNC to SMB .............................................

 

 

HP 85680-60093

BNCTee ............................................................

 

 

HP1250-0781