ADJUSTMENTS

/--

5-22. 3 dB BANDWIDTH ADJUSTMENTS(SERIAL PREFIX 1909A)

REFERENCE:

A9 Schematic

DESCRIPTION:

The 3-dB bandwidths for the 3 MHz, 1 MHz and 300 kHz RESOLUTION BW settings are adjusted using the CAL OUTPUT as the signal source. The 3-dB bandwidths for the 10 kHz, 3 kHz, and 1 kHz RESOLUTION BW settings are adjusted by injecting a stable 321.4 MHz signal into the third converter of the spectrum analyzer.

DISPLAY

SIGNAL GENERATOR

FREQUENCY COUNTER

 

 

 

 

13(2I3

EXTENDERCABLE

-),

11111, ,111,

SPECTRUM

ASSEMBLY

 

1) (

ANALYZER

 

 

3 ,

OUTPUT

FIGURE5-8.3dBBANDWIDTH ADJUSTMENTTESTSETUP

PROCEDURE:

1.Set spectrum analyzer controls asfollows:

FREQUENCY BAND GHz

.................................................. .01- 3

TUNING .................................................................

35 MHz

FREQ SPAN/DIV .........................................................

200 kHz

RESOLUTION BW .........................................................

1MHz

INPUTAmEN .............................................................

20dB

REF LEVEL dBm ...............................................................

0

Amplitude Scale ..............................................................

LIN

SWEEP TIME/DIV .........................................................

1msec

SWEEPTRIGGER .....................................................

FREERUN

VIDEOFILTER .............................................................

MIN

2.Connect equipment as shown in Figure 5-8 except for signal input to AlOW1. Connect CAL OUTPUT to spectrum analyzer INPUT 5(X1.