MODEL 8559A

PERFORMANCETESTS

PERFORMANCE TESTS

4-15.RESOLUTIONBANDWIDTH ACCURACY (Cont'd)

PROCEDURE:

Part of this test must be performed with power supplied to the instrument and with protective covers removed. The test should be performed only by sewice-trainedpersonnelwho are aware of the hazards involved.

1. Set equipment controls as follows:

 

 

Spectrum Analyzer:

 

 

FREQUENCYBANDGHz

..................................................

.01-3

TUNING ..............................................................

0.035GHz

FREQSPAN/DIV ...............................................................

0

RESOLUTIONBW .........................................................

3MHz

INPUTATTEN .............................................................

10dB

REFERENCELEVEL .......................................................

OdBm

REFLEVELFINE ..............................................................

0

AmplitudeScale ..............................................................

LIN

SWEEPTIME/DIV ........................................................

AUTO

SWEEPTRIGGER .....................................................

FREERUN

ALTIF .....................................................................

OFF

SIG IDENT .................................................................

OFF

BLCLIP ....................................................................

OFF

VIDEOFILTER .............................................................

OFF

Signal Generator:

 

 

COUNTERMODE ..............................................

INT,EXPANDXlO

AM ........................................................................

OFF

FM ........................................................................

OFF

FREQUENCYTUNE ......................................................

35MHz

RF ..........................................................................

ON

OUTPUTLEVEL ..........................................................

OdBm

2.Connect equipment as shown in Figure 4-9.

3.Adjust spectrum analyzer TUNING control to locate peak of 35-MHz signal on CRT. Reduce signal generator output if necessary.

NOTE

If necessary, select 10 dBlDlV to locate signal, then switch to LIN.

4.Adjust signal generator OUTPUT LEVEL to position trace at 7.1 divisions above graticule baseline.