MODEL 8559A

PERFORMANCETESTS

PERFORMANCE TESTS4-23. REFERENCE LEVEL ACCURACY (Cont'd)Vernier Accuracy

9.Replace 10-dB step attenuator with 1-dB step attenuator. Set spectrum analyzer controls as follows:

FREQUENCY BAND GHz

.................................................. .01- 3

TUNING ..............................................................

0.030GHz

FREQSPAN/DIV ..........................................................

50kHz

RESOLUTION BW ..............................................

300 kHz, uncoupled

INPUTATTEN ..............................................................

OdB

REFERENCE LEVEL ....................................................

- 10 dBm

REFLEVELFINE ..............................................................

0

AmplitudeScale .........................................................

1 dB/DIV

SWEEPTIME/DIV ........................................................

AUTO

SWEEPTRIGGER .....................................................

FREERUN

ALT IF .....................................................................

OFF

SIGIDENT .................................................................

OFF

BLCLIP ....................................................................

OFF

VIDEOFILTER ..........................................................

2o'clock

10.Set step attenuator to 0 dB. Center signal on CRT and adjust signal generator OUTPUT LEVEL control to position peak of signal 6 divisions above graticule baseline. Set step attenuator and spectrum analyzer REFERENCE LEVEL FINE control to settings indicated in Table 4-13. Record deviation of signal peak from 6th division for each setting.

11.To compute Corrected Deviation, add Step Attenuator Error to Deviation from 6th Division for each setting. Corrected Deviation should not exceed k 0.5 dB for each setting. Record maximum value.

TABLE 4-13.VERNIER ACCURACY

 

 

-- -

 

- --

 

 

 

 

Step Attenuator

 

REFERENCE

Deviation from

Step Attenuator

Corrected

Setting

 

LEVEL FINE

6th Division

Error (Calibration)*

Deviation

(dB)

 

Setting

(dB)

(dB)

(dB)

 

 

 

 

 

 

 

 

 

0(Ref)