PERFORMANCETESTS

MODEL8559A

PERFORMANCE TESTS

,

4-26. DISPLAY FIDELITY

SPECIFICATION:

Log Incremental Accuracy: f0.1 dB per dB from Reference Level

Log Maximum Cumulative Error: <+_ 1.5 dB over entire 70-dB range

Linear Accuracy; t-3% of Reference Level

DESCRIPTION:

The amplitude of the log display amplifier is tested by connecting a DVM to the display (AUX A) VERTICAL OUTPUT connector. A wide resolution bandwidth setting is selected so the signal appears as a straight horizontal line on the CRT The DVM is used to provide good resolution when checking for +1 dB per 10 dB step (0.1 dB/dB).

SIONAL GENERATOR

D I O I T A L VOLTMETER

SPECTRL

ADAPTER 1

 

STEP

ADAPTER A

 

 

ATTENUATOR

 

 

 

-

 

 

 

 

 

 

 

 

 

FIGURE 4-22.DISPLAYFIDELITY TEST SETUP

 

 

EQUIPMENT

 

 

 

 

 

SignalGenerator .........................................................

 

 

 

HP8640B

Digitalvoltmeter ........................................................

 

 

 

HP3556A

Step Attenuator (10 dB/step) ......................................

 

 

HP 355D, Opt. H82

Adapter, Type N (m) to BNC (f)

(2 required) ...............................

HP 1250-0780

Cable, BNC to Banana Plug ..............................................

 

 

 

HP 11001A

PROCEDURE:

Log Display Accuracy

1.Set equipment controls as follows: Spectrum Analyzer:

FREQUENCYBANDGHz

.................................................. .01-3

TUNING ..............................................................

0.030GHz