ADJUSTMENTS

MODEL 8559A

ADJUSTMENTS

5-25. FIRST CONVERTER ADJUSTMENTS

REFERENCE:

A3, A4, AS, A6, and A7 Schematics

DESCRIPTION:

The First LO (A6 YTO) is adjusted by monitoring the YTO output at the RF input connector (LO emission) and the tuning voltage (TUNE) output of the A7 Frequency Control board, and adjusting the YTO low-end frequency for 3 GHz at OV tuning voltage and 6 GHz at - 10Vtuning voltage.

D l SPLAY

n

EXTENDER

CABLE

ASSEMBLY

L

MICROWAVE

COUNTER

DIGITA L VOLTMETER

 

 

CABLE

 

 

ASSEMBLY

 

 

A7TP4

A 1 6 T P l

 

TUNE

LOCATED THROUGH

 

\

HOLE I N S I D E

 

: z SPECTRUM

FRAME HERE --C' O

 

.@.@@@ ANALYZER

 

..no

O...Q

CABLE

R Q 2 1 4-l-U

COMB

QENERATOR

/\

1111

OUTPUT

I

FIGURE 5-11. FIRST CONVERTER ADJUSTMENTSTEST SETUP

 

EQUIPMENT:

 

FrequencyCounter .......................................................

HP5342A

Digital Voltmeter ........................................................

H P 3456A

Comb Generator ........................................................

H P 8406A

Cable, BNC (m) to Banana Plug ...........................................

H P 10111A

Extender Cable Assembly ..............................................

H P SO6O-0303

Adapter, BNC (f) to Alligator Clips ......................................

H P 8120-1292

Cable Assembly, RG-214/U, Type N Connectors .............................

H P 11500A

PROCEDURE:

1.Allow one-half hour warmup time of equipment with spectrum analyzer connected to mainframe with extender cable.