SERVICE

MODEL 8559A

PROCEDURE:

 

Set H P 8559A Spectrum Analyzer controls as follows:

 

TUNING ......................................................

.035 GHz

FREQ SPAN/DIV .................................................

1MHz

RESOLUTION BW ...............................................

300 kHz

INPUT ATTEN .....................................................

0 dB

REFERENCE LEVEL ...........................................

- 10 dBm

REFERENCE LEVEL FINE .............................................

0

Amplitude Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10dB/DIV

SWEEP TIME/DIV ...............................................

AUTO

SWEEP TRIGGER ............................................

FREE RUN

VIDEOFILTER .....................................................

OFF

BL CLIP ...........................................................

OFF

SIGIDENT ........................................................

OFF

ALT IF ............................................................

OFF

Connect equipment as shown. Set signal generator for a 35 MHz, - 10 dBm output signal. Center the Cal signal on the display and adjust for top graticule.

Using board extenders when necessary, check voltages and waveforms indicated on schematic diagrams. Trigger oscilloscope on negative transition of AUX B PENLIFT/BLANKING signal from rear of display mainframe.

To measure RF power levels, set RESOLUTION BW control to 3 MHz and FREQ SPAN/DIV to 0 (zero span). The first LO is not swept in zero span, allowing signal levels to be checked with a second spectrum analyzer (use adapter cables as necessary). DO NOT use a power meter (harmonics and LO signals will contribute to give erroneous levels).

FIGURE 8-2.CONDITIONS FOR SCHEMATIC DIAGRAM MEASUREMENTS (2 OF 2)