Message | Description |
Message | Feeder self test failed |
Action | The |
| sequence on the scanner. If the problem persists, call Service. |
|
|
Message | Front image cal unsuccessful |
Action | The imaging calibration for the front camera was unsuccessful. Clean the |
| imaging guides and retry calibration. See Chapter 5, Maintenance, “Cleaning the |
| imaging guides” for cleaning procedures. |
|
|
Message | Image address field overflow |
Action | One of the image address fields has exceeded its maximum configured count. |
| Check to make sure that a level change was not missed or reconfigure the |
| scanner to allow a larger value for the overflowed field. |
|
|
Message | Image buffer almost full |
Action | The feeder has stopped and the transport is still running because the image buffer |
| is almost full. Processing will resume when the buffer is less full. |
|
|
Message | Image buffer full |
Action | The image buffer overflowed. |
| One or more documents that passed through the transport were not scanned. |
| Refeed all documents for which images were not obtained. |
|
|
Message | Images in buffer |
Action | You tried to enable the scanner when images were still in the buffer. Not all |
| images were retrieved. Retrieve the remaining images or clear the buffer. When |
| the buffer is cleared, you may enable the scanner. |
|
|
Message | Interlocks are closed |
Action | Informational message. All doors are closed. |
|
|
Message | Interlocks are open |
Action | One of the doors is not closed or an error has not been cleared since the door(s) |
| was last opened. Check to see that all doors are closed and clear the error by |
| pressing the Return arrow (left button) on the Control Panel. |
|
|
Message | Invalid IA level |
Action | The image address level you tried to enter is not valid. |
| address level. For more information about image addressing, see Kodak |
| publication, Kodak i800 Series Scanners, Image Processing Guide, |
|
|
Message | IPB needs firmware |
Action | There is no firmware in an image processing board. Firmware needs to be |
| installed in all image processing boards before scanning is possible. Call Service. |
|
|
Message | Lamps not ready |
Action | Calibration and/or scanning was attempted before the lamps were warmed up. |
| See Chapter 3, Operator Control Panel Function’s for calibration guidelines and |
| procedures. |
|
|
Message | Lamps ready for scanning |
Action | Although lamps have warmed up sufficiently to allow scanning, image quality will |
| be best after the lamps have stabilized after approximately 10 minutes. |
|
|
|