5. Turn the test strip over and check to see that the confirmation dot is completely blue. If it is not, discard the test strip.

Repeat the test with a new test strip, applying a larger drop of control solution to the centre of the pink test square.

C H E C K I N G

6. Firmly insert the test strip, confirma- tion dot side down, into the test strip holder until

it stops.

CAUTION: If you do not firmly insert the test strip into the meter until it stops, you may get an inaccurately low result.

IMPORTANT: Immediately insert the test strip after applying the control solution. If you do not insert the test strip within 2 min- utes, discard the test strip and start over with a new test strip.