12.Confirm that the meter is still on. Firmly insert the test strip, tapered- end first, confir- mation dot side down, into the test strip holder until it stops.

CAUTION: If you do not firmly insert the test strip into the meter until it stops, the test may start but you may get an inac- curately low result. Do not remove test strip until countdown is complete.

IMPORTANT: Immediately insert the test strip after applying the blood. If you do not insert the test strip within 2 min- utes, discard the test strip and start over with a new test strip.

31

A flashing clock symbol appears while the meter is reading your result. Your glucose result appears in about 30 seconds. Do not

remove strip until countdown is com- plete. The wait may be shorter or longer depending upon the current temperature.

The test result is displayed alternating with the time, then with the date. The meter displays results between

0-27.8 mmol/L (0 and 500 mg/dL). If the result is above 27.8 mmol/L (500 mg/dL) the meter displays HI in virtually all cases.

(Examples)

T E S T I N G

Page 40
Image 40
Lifescan SPD2410BD manual Examples