Coding the Meter

A code number is printed on each test strip vial label. The code numbers (ranging from 1 to 21) are used to calibrate the meter.

Each time you test, check that the code number on the meter display matches the code number on the test strip vial label.

C AU T I O N : If the code number on the meter display does NOT match the code number on the test strip vial label, you may get inaccurate results.

1. Press the blue power button to turn on the meter.

Make sure there is not a test strip in the meter when you turn it on.

Display Check

(Example)

The display check appears briefly. Carefully check that all symbols appear. The result of your last blood glucose or control solution test appears.

If you are turning on the meter for the first time, –––

appears on the display.

Next, a code number appears for a few seconds.

C O D I N G