TROUBLESHOOTING

NOTE: If the actions listed below do not resolve the error, call the OneTouch® Customer Care Line at 1 800 663-5521.

ERROR POTENTIAL PROBLEM WHAT TO DO

Er 1

The meter detected

Professional Use:

continued

a problem with the

Test the sample with an alternate method.

colour of the test

 

While awaiting results:

 

 

 

strip.

 

 

 

Compare the colour of the test strip confirmation dot

 

 

 

 

 

 

to the Colour Chart on the test strip vial label. If the

 

 

 

glucose level is very low (2.8 mmol/L) or very high

 

 

 

(19.4 mmol/L), take appropriate precautions

 

 

 

immediately.

 

 

 

Before retesting verify that the sample hematocrit is

 

 

 

within range.

 

 

If these actions do not resolve the error, call the

 

 

 

OneTouch® Customer Care Line at 1 800 663–5521.

 

 

 

 

 

The meter detected

1.

Repeat the test with a new test strip and:

 

a problem with the

 

Check the test strip expiration date. Do not use

 

 

 

test strip.

test strips beyond the expiration or discard

 

 

date.