blood glucose is very low

(2.8 mmol/L), conduct a Control Solution Test (see “Checking The System”). If the Control Solution test indicates that your meter and test strips are working together properly, repeat your blood glucose test with a new test strip. If you continue to receive a very low blood glucose test result, the low glucose result is likely to be correct and you should follow your Healthcare Professional’s advice for treating low glucose levels. If the Control Solution test is out of range, call the OneTouch® Customer Care Line at 1 800 663-5521.

IMPORTANT: When proper procedures are followed, the meter will display “HI” in virtu- ally all cases when your result is above

27.8 mmol/L (500 mg/dL). The confirmation dot colour will also be darker than the

“19.4 mmol/L” sample colour dot on the test strip vial label. This indicates very high blood glucose (severe hyperglycemia). Follow the advice of your Healthcare Professional. If you have never experienced blood glucose levels this high or have not been instructed in how to deal with them by your Healthcare Professional, you should contact your Healthcare Professional at once.

How to Check if Your

Meter is Working Correctly

The easiest way to check if your meter and test strip are working together properly is to run a test with OneTouch® SureStep® Glucose Control Solution. See Checking the System for more information.

R E S U L T S

37