◆IMPORTANT Test Strip Information
For detailed information regarding SureStepPro Test Strips, read the package insert.
•Use only SureStepProTest Strips with the SureStepPro Blood Glucose Bedside Unit.
•SureStepPro Test Strips are sensitive to heat, light, and moisture. Keep them tightly sealed in their original bottle, and store in a cool, dry place below 30°C (86°F). Do not refrigerate, freeze, or place in direct heat or sunlight.
•A lot number and CTL (control) code are printed on the test strip bottle label. (The CTL code encodes the test strip expiration date, control solution ranges, and linearity reference values.) When entering this information, always get the lot number and CTL code from the test strips currently in use.
•Do not use the test strips after the expiration date printed on the bottle label.
•Discard any unused test strips 4 months after opening. Write the opened date on the bottle when you first open it.
•Replace the test strip bottle cap immediately after removing a test strip and close the cap tightly.
•Do not transfer test strips to a new bottle or any other container.