Lifescan SureStep Pro manual Do not remove the test strip until the countdown is complete

Models: SureStep Pro

1 117
Download 117 pages 28.59 Kb
Page 86
Image 86

Special Tests

8Firmly insert the test strip all the way into the test strip holder until it stops (the confirmation dot should be facing down).

IMPORTANT

If you fail to completely insert the test strip, the test may start; however, you may receive an inaccurate, low result.

IMPORTANT

You have up to 2 minutes to insert the test strip after the solution is applied. If you insert the test strip after 2 minutes, you may get an inaccurate result or an error message. Discard the test strip and repeat the test with a new test strip.

The test result appears in approximately 30 seconds.

IMPORTANT

Do not remove the test strip until the countdown is complete.

Unkn A Result

 

 

mg/dL

Menu

Enter

Ok