Unexpected Messages and Observations
MESSAGE | POSSIBLE | SOLUTION | |
CAUSE | |||
|
| ||
|
|
| |
Apply Blood | Bedside unit is | Obtain a drop of capillary | |
| ready for blood | blood and apply it to center | |
| application. | of pink test square. | |
|
| • If white pad becomes satu- | |
|
| rated, you have applied | |
|
| too much blood. Repeat | |
|
| application with new test | |
|
| strip. | |
|
| • If white patches or streaks | |
|
| appear on confirmation | |
|
| dot, you have not applied | |
|
| enough blood. Repeat | |
|
| application with new test | |
|
| strip. | |
|
|
| |
Bedside Unit | Bedside unit | Contact system | |
Disabled | deleted from | administrator for current | |
| system or | status of bedside unit. | |
| unassigned. |
| |
|
|
| |
Bedside Unit | Initial data transfer | Transfer data. | |
Needs | is required. Or, |
| |
Configuration | configuration is |
| |
| corrupted. |
| |
|
|
|
If you cannot resolve a problem, call LifeScan Healthcare Professional Line: