IMPORTANT

If you fail to completely insert the test strip, the test may start; however, you may receive an out-of-range result.

IMPORTANT

You have up to 2 minutes to insert the test strip after the linearity solution is applied. If you insert the test strip after 2 minutes, you may get an out-of-range result or an error message. Discard the test strip and repeat the test with a new test strip.

The test result appears in approximately 30 seconds. The time may vary depending on the temperature and glucose level.

IMPORTANT

Do not remove the test strip until the countdown is complete.

NOTE

The bedside unit can display linearity results up to 766 mg/dL.