MESSAGE

POSSIBLE

SOLUTION

CAUSE

 

 

 

 

 

Data

Less than 100 test

Transfer data to free memory

Memory Low

records can be

space.

 

generated.

 

 

 

 

Data Transfer

Bedside unit/

Retry data transfer.

Error

workstation

 

 

communication

 

 

error.

 

 

 

 

Enter

High, low, or

Check ranges. Make sure min

Narrower

normal control

value is less than max.

Range(s) or

range error.

 

Min<Max

Min>Max.

 

 

 

 

Insert Strip

Test strip not fully

Remove test strip. When

appears after

inserted into test

“Insert Strip” appears, firmly

you insert test

strip holder.

insert test strip until it stops.

strip.

 

 

 

 

 

Internal

Faulty internal

Retry operation. If problem

Communica-

communications.

persists, do not use bedside

tion Error

 

unit. Contact LifeScan.

 

 

 

Internal Error

System Error.

Turn off bedside unit.

Power Down

 

Contact LifeScan.

 

 

 

Invalid

Conflicting option

Contact system

Configuration

set, or

administrator to review

 

configuration

configuration settings.

 

corrupted.

Transfer data.

 

 

 

Invalid

Value

Re-enter correct control

Control

>500 mg/dL was

value.

Value

entered.

 

 

 

 

Invalid CTL

Expiration date not

Check CTL code on control

Code

correct, or CTL code

solution vial and re-enter it.

 

format invalid.

 

 

 

 

Invalid

Invalid month or

Check expiration date and

Expiration

year. Or, date

re-enter it.

Date

entered already

 

 

expired.

 

 

 

 

If you cannot resolve a problem, call LifeScan Healthcare Professional Line:

1 800 524-7226 (USA)

1 888 353-0800 (Canada)

87

Page 99
Image 99