▲CAUTION
Do not apply additional blood to the test strip or you may get an inaccurate result.
If white patches or streaks continue to appear after you have repeated the test with a new test strip and a larger sample volume, call the LifeScan Healthcare Professional Line at
1 800
9Firmly insert the test strip all the way into the test strip holder until it stops (the confirmation dot should be facing down).
▲CAUTION
If you fail to completely insert the test strip, the test may start; however, you may receive an inaccurate, low result.
▲CAUTION
You have up to 2 minutes to insert the test strip after the blood is applied. If you insert the test strip after 2 minutes, you may get an inaccurate result or an error message. Discard the test strip and repeat the test with a new test strip.
37