Quality Control
Configuration Option | Meaning |
|
|
|
|
Glucose units | The units selected for |
❍ mg/dL | displaying and reporting |
❍ mmol/L | control and blood glucose |
| results. |
|
|
Required QC tests | Three glucose controls are |
❑ High | available. Any combination |
❑ Low | may be selected. |
❑ Normal |
|
|
|
Control solution | The passing range for each |
ranges (high, low, | control solution level has |
normal) for specified | been defined. This range |
test strip | may be the same as or |
| different from the range |
| found on the test strip |
| bottle label. |
|
|
Result Display | Determines how quality |
❍ Numeric Value | control test results are |
❍ Pass/Fail Only | displayed. |
(Pass/Fail available for |
|
SureStepPro 2.0 only) |
|
|
|
QC Enforcement | An option to ensure QC is |
❍ Warning Only | successfully completed as |
❍ Bedside Unit | scheduled. Warning Only |
lockout | displays a message but |
❑ Operator | allows you to perform a |
Lockout | patient test. Bedside Unit |
| lockout displays the same |
| message but does not allow |
| you to perform a patient |
| test. Operator Lockout |
| prevents the operator who |
| has not successfully per- |
(Time period for | formed QC testing within |
Operator Lockout is | the designated time period |
24 hours for | (1 to 365 days) from |
SureStepPro 1.2) | performing a patient test. |
|
|
95