2Wipe the test strip holder cover and base with a cotton swab or soft cloth moistened with a 10% bleach solution. Be sure to thoroughly wipe the area on the inside cover. Clean both sides of the base. Follow with a swab moistened with water to remove residual bleach.

CAUTION

Bleach residue left on the test strip holder may lead to an error message or an inaccurate, high result.

3Dry the test strip holder with a soft cloth or lint-free tissue. Close the holder and set it aside.

4Using a cotton swab or soft cloth moistened with a 10% bleach solution, wipe the lens area and contact points. Wipe this area even if it does not appear to be dirty. Use a swab or cloth moistened with water to remove residual bleach. Be careful not to scratch the lens area.

contact points

lens area

55

Page 67
Image 67
Lifescan SureStep Pro manual Contact points Lens area