Troubleshooting

ERROR

POSSIBLE

SOLUTION

CAUSE

 

 

 

 

 

Error 4

The bedside unit

1.Repeat test with new test

Temperature

detected a

strip. Make sure bedside

Range

temperature

unit and test strip are at

 

problem.

room temperature.

 

 

 

• Do not apply heat or

 

 

cold to unit (operating

 

 

temperature is 18°–30°C).

 

 

2.If you continue to get any

 

 

error message, obtain a

 

 

laboratory result, and

 

 

• Compare color of test

 

 

strip confirmation dot to

 

 

Color Chart on test strip

 

 

bottle to see if glucose

 

 

level is very high or very

 

 

low.

 

 

• Take appropriate steps

 

 

to monitor patient and

 

 

provide medical care while

 

 

awaiting laboratory result.

 

 

3.Contact LifeScan for help.

 

 

 

If you cannot resolve a problem, call LifeScan Healthcare Professional Line:

1 800 524-7226 (USA)

1 888 353-0800 (Canada)

82

Page 94
Image 94
Lifescan SureStep Pro manual Detected a