Troubleshooting
ERROR | POSSIBLE | SOLUTION | |
CAUSE | |||
|
| ||
|
|
| |
Error 4 | The bedside unit | 1.Repeat test with new test | |
Temperature | detected a | strip. Make sure bedside | |
Range | temperature | unit and test strip are at | |
| problem. | room temperature. | |
| |||
|
| • Do not apply heat or | |
|
| cold to unit (operating | |
|
| temperature is | |
|
| 2.If you continue to get any | |
|
| error message, obtain a | |
|
| laboratory result, and | |
|
| • Compare color of test | |
|
| strip confirmation dot to | |
|
| Color Chart on test strip | |
|
| bottle to see if glucose | |
|
| level is very high or very | |
|
| low. | |
|
| • Take appropriate steps | |
|
| to monitor patient and | |
|
| provide medical care while | |
|
| awaiting laboratory result. | |
|
| 3.Contact LifeScan for help. | |
|
|
|
If you cannot resolve a problem, call LifeScan Healthcare Professional Line: