Bedside Unit Configuration Options
Operator and IDs
Configuration Option | Meaning |
|
|
|
|
❑ ID required for all | If selected, all operators are |
tests | required to enter their IDs |
❑ Maintain ID | for each test. |
for ____ | If Maintain ID is selected, |
minutes after | the last ID entered remains |
power off | in memory for 1 to 60 |
| minutes from the time the |
| unit is turned off until it is |
| turned on again. Operators |
| need only confirm their ID |
| if it appears at the operator |
| ID screen. |
|
|
❑ Patient ID required | If selected, a patient ID is |
for all tests | required. |
|
|
ID number string | A string format was |
format (patient and | defined for all IDs entered |
operator) | at the bedside unit. The |
| string defines the number |
| and type of characters |
| allowed (for example, five |
| alphanumeric characters). |
|
|
Operator | The period is the frequency |
Recertification | at which each operator |
Period, ___months | must be |
and test settings | settings define the number |
| of patient tests as well as |
(test settings | low, normal, and high QC |
available for | tests an operator must run |
SureStepPro 2.0 only) | within the |
| period to be eligible for |
| |
|
|
94