Troubleshooting

ERROR

POSSIBLE

SOLUTION

CAUSE

 

 

 

 

 

Error 6

The bedside unit

1.Wipe contact points in lens

System Error

detected a problem

area with cloth moistened

 

with the electronics.

with water to remove

 

 

 

residual bleach. Dry

 

 

thoroughly, but

 

 

NOTE: Before cleaning

 

 

contact points, compare

 

 

color of test strip confirma-

 

 

tion dot to Color Chart on

 

 

test strip bottle. If glucose

 

 

level is very high or very low,

 

 

take appropriate steps to

 

 

monitor patient and provide

 

 

medical care.

 

 

• Repeat test with new

 

 

test strip.

 

 

• Make sure test strip lot

 

 

number (and code)

 

 

entered at bedside unit

 

 

matches number on test

 

 

strip bottle.

 

 

2.If you continue to get any

 

 

error message, obtain a

 

 

laboratory result. Take

 

 

appropriate steps to

 

 

monitor patient and

 

 

provide medical care while

 

 

awaiting laboratory result.

 

 

3.Contact LifeScan for help.

 

 

 