Bedside Unit Configuration Options
Configuration Option | Meaning |
|
|
|
|
Critical Warning | Notes that appear at the |
Messages | patient test result screen |
Lower Limit ____ | when the test result falls |
Upper Limit ____ | outside the critical limits. |
| The notes are intended to |
| offer suggestions for |
| abnormal results. |
| The upper and lower limits |
| of the range for patient |
| results as defined by your |
| institution. |
|
|
Patient Message | If a patient test result falls |
Entry Required (for | outside the critical limits, |
results outside | you may be required to |
critical limits) | select a comment which |
❍ Yes | corresponds to the patient’s |
❍ No | current situation. |
|
|
Bedside Unit
Configuration Option | Meaning |
|
|
|
|
❑ BU lockout if no | If selected, the bedside |
upload in ____ | unit cannot be used for |
days | testing if a data transfer |
| session has not been |
| performed within the |
| specified number of days |
| (maximum of 60 days). |
|
|
❑ Bedside unit | If selected, optional |
beeper on | bedside unit beeps are |
| enabled. |
|
|
❑ Enable barcode | If selected, the bedside |
scanner | unit barcode scanner is |
| enabled. If not selected, |
| you must enter all ID and |
| lot number information |
| manually. |
|
|
92