Bedside Unit Configuration Options

Configuration Option

Meaning

 

 

 

 

Critical Warning

Notes that appear at the

Messages

patient test result screen

Lower Limit ____

when the test result falls

Upper Limit ____

outside the critical limits.

 

The notes are intended to

 

offer suggestions for

 

abnormal results.

 

The upper and lower limits

 

of the range for patient

 

results as defined by your

 

institution.

 

 

Patient Message

If a patient test result falls

Entry Required (for

outside the critical limits,

results outside

you may be required to

critical limits)

select a comment which

Yes

corresponds to the patient’s

No

current situation.

 

 

Bedside Unit

Configuration Option

Meaning

 

 

 

 

BU lockout if no

If selected, the bedside

upload in ____

unit cannot be used for

days

testing if a data transfer

 

session has not been

 

performed within the

 

specified number of days

 

(maximum of 60 days).

 

 

Bedside unit

If selected, optional

beeper on

bedside unit beeps are

 

enabled.

 

 

Enable barcode

If selected, the bedside

scanner

unit barcode scanner is

 

enabled. If not selected,

 

you must enter all ID and

 

lot number information

 

manually.

 

 

92