Setting

DATE

SENSOR SERIAL NUMBER FINAL ASSEMBLY NUMBER TV1 Units

TV1 LRV

TV! URV

Description

"customer input data via HART" "customer input data via HART" "customer input data via HART"

UNDEFINED

0[switch position (0)E, TV1]

1.0[switch position (0)F, TV1]

Interconnection

The Upper Sensor Limit (USL) and Lower Sensor Limit (LSL) are set to 3300 and 1.666 pF respectively, and the following conditions apply: the Upper Range Value (URV) and Lower Range Value (LRV) should be within the USL to LSL range, but can be set anywhere within that range.

An interruption of the measuring connection will be detected: a loose or interrupted connection results in up to 0.5 pF capacitance, which is below the adjusted LSL and thus signals a FAULT condition.

7ML19985GE01

SITRANS LC 500 – INSTRUCTION MANUAL

Page 35