Siemens LC 500 Additional Universal Command Specifications, MultiDrop operation, Burst mode

Models: Sitrans

1 122
Download 122 pages 51.76 Kb
Page 86
Image 86
MultiDrop operation

Appendix C

MultiDrop operation

The SITRANS LC 500 transmitter supports MultiDrop Operation.

Burst mode

The SITRANS LC 500 transmitter does not support Burst Mode.

Units conversions

The Transmitter Variable #0 Units are in pF and cannot be changed.

The Transmitter Variable #1 Values may be set to any Units and Value with Command #140. The Transmitter Variable Range Values may be read at any time with Command #141.

The value returned as Secondary Variable (S.V.) is the result of the following calculation: TV1 = TV0 Dynamic Range Value in percent x ({TV#1}URV – {TV#1}LRV) + {TV#1}LRV.

This method provides a means of converting TV#0, which is always in pF, to alternative units (level or volume).

Additional Universal Command Specifications

For a document listing the additional Universal Command Specifications, please contact Technical Publications at techpubs@siemens-milltronics.com