Troubleshooting: SITRANS LC 500

If you are unable to change settings:

check that keylock level (menu 1F) is set to 0: the display should read PL = 0

check that menu 01 is set appropriately: if Pv = 1, changes can only be made via

HART

If you can change settings:

reset menu 12 to factory settings: press both buttons, and the display should read FAC A

If the LCD displays a negative reading, typically around minus 300 pF, this often indicates a short circuit in the probe assembly:

check the enclosure and make sure no water has got in

check that all the connections in the probe assembly are solid