Image 62
Range: 1 to 10,000

Appendix A

Damping

Damping slows the measurement response to a change in level, and is used to stabilize the reading1. The Damping Value is not in seconds but is a factor that controls the rate of change for the dynamic value of the TV currently selected.

The increment/decrement step size is subject to the setting on Menu 09.

Menu

Affected

Rotary

Left

 

 

Switch

Description

Values

Item

by:

Position

Arrow

 

 

0A

01, 09

A

Off

Damping

Range: 1 to 10,000

 

 

Factory setting

1.00

 

 

 

 

 

 

 

 

 

 

1.Set the rotary switch to A.

2.Press the RED (+) or the BLUE (–) button to alter the value between 1 and 10,000. or: Press and hold a button to start a repeat function,

or: Press and hold both buttons simultaneously to reset the value back to 1.00.

Lower Sensor Limit

The Lower Sensor Limit (LSL) is the lower of two limit settings. Whenever the PV value (Menu level 0) drops below the Lower Sensor Limit, the measurement is considered at fault and the LCD displays ooL, alternating with PV.

If the display mode is in %, this selection is disabled and the LCD displays - - - -.

The transmitter variable on which this menu selection operates is chosen in Menu 01. The increment/decrement step size is subject to the setting from Menu 09.

Menu

Affected

Rotary

Left

 

 

Switch

Description

Values

Item

by:

Position

Arrow

 

 

0B

01, 09

B

Off

Lower Sensor Limit

Range 0 to 3300

 

 

Factory setting

1.66

 

 

 

 

 

 

 

 

 

 

1.Set the rotary switch to B.

2.Press the RED (+) or BLUE (–) buttons to alter this value. or: Press and hold a button to start a repeat function,

or: Press and hold both buttons simultaneously to take the current PV reading as the new setting.

1.For example, in an application with an agitated surface.

Page 56

SITRANS LC 500 – INSTRUCTION MANUAL

7ML19985GE01

Page 62