Appendix A

The transmitter variable on which this selection is based is chosen in Menu 01. The increment/decrement step size is set at Menu 09.

Menu

Affected

Rotary

Left

 

 

 

 

Switch

Mode

 

Description

Values

Item

by:

Arrow

 

Position

 

 

 

 

 

 

 

 

2-state

 

Delta Range Setting

Span

 

 

 

 

 

(enabled)

setting

 

 

 

 

 

 

0D

01, 07, 09

D

Off

 

 

 

 

 

 

 

Display

 

 

 

 

Analog

*

 

shows

 

 

 

 

 

 

 

- - - -

 

 

 

 

 

 

 

 

Lower Range Value

Lower Range Value (LRV) is the setting for 0% of the operating range, in most cases an empty vessel/tank. If the display mode is in % this selection is disabled and the LCD displays - - - -.

The transmitter variable on which this menu selection operates is chosen in Menu 01. For TV0 the LRV is in most cases Factory Set to the probe capacitance in free air.

The increment/decrement step size is subject to the setting from Menu 09.

Menu

Affected

Rotary

Left

 

 

 

Switch

Mode

Description

Values

Item

by:

Arrow

Position

 

 

 

 

 

 

 

 

Lower Range

Range: 0.00 to 3300

 

 

 

 

 

Value

 

 

 

 

 

 

 

 

 

 

 

 

 

0E

01, 09, 0B, 0C

E

Off

Analog

Factory setting

Probe capacitance in

air

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Display percent

Display shows

 

 

 

 

 

 

- - - -

 

 

 

 

 

 

 

1.Set the rotary switch to E.

2.Press and hold both buttons simultaneously to take the current PV reading as the new setting.

or: Press the RED (+) or BLUE (–) button to step the value up or down.

or: Press and hold a button for a prolonged time to start a repeat function.

When the new setting exceeds that of the Limit Settings (Menu 0C and 0B), the new value is rejected and the previous value remains unchanged.

Page 58

SITRANS LC 500 – INSTRUCTION MANUAL

7ML19985GE01