Appendix A

Analog Output Signalling (proportional or 2-state): menu level 0

Analog mode (the loop-current) can provide either:

a 4 to 20 / 20 to 4 mA output proportional to the percent of the range

or

a 4 or 20 / 20 or 4 mA output, when 2-state mode is selected

Note: To set values for Upper and Lower Threshold Delay, and Upper and Lower Threshold Setting (2-state mode), the loop-current menu (07) must be in 2-state mode. When the loop-current control is in analog mode, the LCD displays only - - - - for these menu selections.

Upper Threshold Delay (2-state mode)

The Upper Threshold Setting controls the Activation delay: the amount of time that has to pass uninterrupted with the probe covered to a level above the Upper Threshold Setting before the timer expires. When the timer expires, the output signal complies with the setting from Menu 08 for a covered probe. Whenever the level drops below the Upper Threshold Setting before the timer expires, the timer is restarted.

As an extra identifier, an upward running A is displayed to the right of the value.

Menu

Affected

Rotary

Left

 

 

Added

 

 

Item

by:

Switch

Arrow

Mode

 

Indicator

Description

Values

 

 

Position

 

 

 

 

 

 

 

 

 

 

2-

 

Upward

Upper

in

 

 

 

 

 

running A at

Threshold

 

 

 

 

state

 

seconds

 

 

 

 

 

right of value

Delay

03

07

3

Off

 

 

 

 

 

 

 

 

 

 

 

 

Display

 

 

 

 

 

 

 

 

 

 

 

 

Analog

*

 

 

shows

 

 

 

 

 

 

 

 

- - - -

 

 

 

 

 

 

 

 

 

1.Set the loop-current control (Menu 07) to 2-state mode.

2.Set the rotary switch to 3.

3.Press the RED (+) or BLUE(–) button to increase or decrease the value. or: Press and hold a button to start a repeat function.

or: Press and hold both buttons simultaneously to toggle the value between minimum (0) and maximum (100).

Lower Threshold Delay (2-state mode)

The Lower Threshold Setting controls the Deactivation delay: the amount of time that has to pass uninterrupted with the probe covered to a level below the Lower Threshold Setting before the timer expires. When the timer expires, the output signal will comply with the setting from Menu 08 for an uncovered probe. Whenever the level rises above the Lower Threshold Setting before the timer expires, the timer is restarted.

As an extra identifier, a downward running A is displayed to the right of the value..

Page 62

SITRANS LC 500 – INSTRUCTION MANUAL

7ML19985GE01