Series 24-HP Instruction ManualAppendix A Specifications

Appendix A Product Specifications

Accuracy

 

 

 

 

 

 

 

 

 

 

 

 

 

Process

 

240 Series In-Line Meters

241 Series Insertion Meters

(1)

Variables

 

 

 

 

 

 

 

 

 

 

Liquids

Gas & Steam

Liquids

Gas &

 

 

 

 

 

 

Steam

 

Mass Flow

 

±1% of rate

±1.5% of

±1.5% of rate

±2% of

 

Rate

 

over a 30:1

rate(2) over a

over a 30:1

rate(2) over

 

 

range(3)

30:1 range(3)

range(3)

a 30:1

 

 

 

 

 

 

range(3)

 

Volumetric

 

±0.7% of

±1% of rate

±1.2% of rate

±1.5% of

Flow Rate

 

rate over a

over a 30:1

over a 30:1

rate over a

 

 

30:1

range(3)

range(3)

30:1

 

 

 

range(3)

 

 

range(3)

 

Temperature

± 2°F

± 2°F

± 2°F

± 2°F

 

 

 

(± 1°C)

(± 1°C)

(± 1°C)

(± 1°C)

Pressure

 

0.4% of

0.4% of

0.4% of

0.4% of

 

 

 

transducer

transducer

transducer full

transducer

 

 

full scale

full scale

scale

full scale

Density

 

0.3% of

0.5% of

0.3% of

0.5% of

 

 

 

reading

reading(2)

reading

reading(2)

Notes:

(1) Accuracies stated are for the total mass flow through the pipe.

 

(2)Over 50 to 100% of the pressure transducer’s full scale.

(3)Nominal rangeability is stated. Precise rangeability depends on fluid and pipe size.

 

Repeatability

 

 

 

Mass Flow Rate: 0.2% of rate.

 

 

 

 

 

 

 

 

 

Volumetric Flow Rate: 0.1% of rate.

 

 

 

 

 

 

 

 

 

Temperature:

± 0.2°F (± 0.1°C).

 

 

 

 

 

 

 

 

 

Pressure: 0.05% of full scale.

 

 

 

 

 

 

 

 

 

Density: 0.1% of reading.

 

 

 

 

 

 

Stability Over 12 Months

Mass Flow Rate: 0.2% of rate maximum.

 

 

 

 

 

 

 

 

Volumetric Flow Rate: Negligible error.

 

 

 

 

 

 

 

 

Temperature: ± 0.1°F (± 0.5°C) maximum.

 

 

 

 

 

 

 

 

Pressure: 0.1% of full scale maximum.

 

 

 

 

 

 

 

 

Density: 0.1% of reading maximum.

 

 

 

 

Response Time

 

 

Adjustable from 1 to 100 seconds.

 

 

 

 

Material Capability

 

 

Series 240 In-Line Flow Meter:

 

 

 

 

 

 

 

 

 

Any gas, liquid or steam compatible with 316L stainless steel, C276

 

 

 

 

 

 

hastelloy or A105 carbon steel. Not recommended for multi-phase

 

 

 

 

 

 

fluids.

 

 

 

 

 

 

 

 

 

 

 

 

 

Series 241 Insertion Flow Meter:

 

 

 

 

 

 

 

 

 

Any gas, liquid or steam compatible with 316L stainless steel. Not

 

 

 

 

 

 

recommended for multi-phase fluids.

 

 

 

 

Flow Rates

 

 

 

Typical mass flow ranges are given in the following table. Precise

 

 

 

 

 

 

flow depends on the fluid and pipe size. 241 insertion meters are

 

 

 

 

 

 

applicable to pipe sizes from 2 inch and above. Consult factory for

 

 

 

 

 

 

sizing program.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Water Minimum and Maximum Flow Rates

 

 

 

 

½-inch

 

¾-inch

 

1-inch

 

1.5-inch

 

2-inch

 

3-inch

 

4-inch

6-inch

8-inch

gpm

1

 

1.3

 

2.2

 

5.5

 

9.2

 

21

 

36

81

142

 

22

 

40

 

67

 

166

 

276

 

618

 

1076

2437

4270

m3/hr

.23

 

.3

 

0.5

 

1.3

 

2.1

 

4.7

 

8.1

18

32

 

5

 

9.1

 

15

 

38

 

63

 

140

 

244

554

970

IM-24-HP

A-1

Page 67
Image 67
Sierra 240-VTP, 241-VTP instruction manual Appendix a Product Specifications